size
Reader: 9.6 x 15.5 x 5.7 cm (width/length/height)
Adapter: 8.6 x 12.8 x 1.7 cm (width/length/height)
-
weight
Lighting unit: 435g
Detection unit: 355g
-
Detection method
Absorbance
-
Detection modeEndpoint, dynamics
-
Microplate type96-well microplate (flat bottom)
-
Measuring range0.0–4.0 OD (optical density)
-
Resolution
0.001 OD (optical density)
-
power supply
USB connection 5 V USB connection 5V
-
Power input
2.5 watts
-
Wavelength filter settings
450 nm, 562 nm, 600 nm, 650 nm, optional customized wavelength
-
Detector
96 Photodiode
-
Linear
405 nm: ≤ 1.5% from 0.0 – 2.0 OD; ≤ 3.0% from 2.0 – 3.0 OD
≥ 450 nm: ≤ 1.0% from 0.0 – 2.0 OD; ≤ 1.5% from 2.0 – 3.0 OD -
Accuracy
405 nm: ≤ 1.5% + 0.010 OD from 0.0 – 2.0 OD; ≤ 3.0% + 0.010 OD from 2.0 – 3.0 OD
≥ 450 nm: ≤ 1.0% + 0.010 OD from 0.0 – 2.0 OD; ≤ 1.5% + 0.010 OD from 2.0 – 3.0 OD -
Repeatability
≤ 0.5 % + 0.005 OD from 0.0–2.0 OD; ≤ 1 % + 0.010 OD from 2.0–3.0 OD